Keysight B2981A (Agilent) FEMTO / PICOAMMETER The Agilent B2981A Femto / Picoammeter is designed with distinctive capabilities and features guaranteeing state-of-the art measurement performance. It offers mimimum cur resolution for meeting all low-level cur measurement needs and its voltage sourcing feature supports resistance measurements. Its unique color LSD provides multiple options for viewing data and front-panel capabilities allow you to capture transient behavior and perform quick statistical analyses without a PC. Features and Specifications of the Keysight B2981A (Agilent) Femto / Picoammeter include:Features:
Graphical capabilities:
- Time domain view to capture transients
- Histogram view to perform quick data analysis
Easy-to-use navigation to select optimal range and aperture
Optional Test Setup Integrity Checker function for noise source isolation
Versatile interface (USB 2.0: LAN, GPIB, LXI Core)
USB (front): store data, save/recall setup information
Free PC control software
Applications:
- Material science (Biomaterials, ceramics, elastomers, films, dielectric materials, electrochemical, ferroelectric materials, graphene, metals, organic materials, nano-materials, polymers, semiconductors, etc.)
- Devices & elecronic components (capacitors, resistors, diodes, sensors, transistors including TFT and CNT, optoelectronics, solar cells, etc.)
- Electronic/non-electronic systems (ion beam, electron beam, sensing systems, particle measurements, embedded precision instruments, etc.)
Time domain view to capture transients
Histogram view to perform quick data analysis
Material science (Biomaterials, ceramics, elastomers, films, dielectric materials, electrochemical, ferroelectric materials, graphene, metals, organic materials, nano-materials, polymers, semiconductors, etc.)
Devices & elecronic components (capacitors, resistors, diodes, sensors, transistors including TFT and CNT, optoelectronics, solar cells, etc.)
Electronic/non-electronic systems (ion beam, electron beam, sensing systems, particle measurements, embedded precision instruments, etc.)
Specifications of the Keysight B2981A (Agilent) Femto / Picoammeter :
Measurement resolution: 6½ digits
Cur measurement: 0.01 fA – 20 mA
Minimum range: 2 pA
Maximum reading rate: 20,000 rdg/s
< 20 µ V burden voltage in its lowest cur range
Voltage measurement:
- Voltage range: 2 V
- Display resolution: 1 µV
- Accuracy ±(% + offset): 0.025 + 40 µV
- Voltage range: 20 V
- Display resolution: 10 µV
- Accuracy ±(% + offset): 0.025 + 400 µV
Input bias cur: < 20 fA
Input impedance: > 200 TΩ, parallel with < 20 pF (non-guarded) or < 2 pF (guarded)
RMS noise: 1.4 µV for 2 V range, 10 s duration, shorted input
Measurement control / navigation:
- Ranging: Automatic or manual
- Aperture time (integration time): 10 µs to 100 PLC
- Offset cancelling: Null, Zero correct
- Digital filter: Medium filter (2R + 1, R = 1 to 15), moving average (1 to 100, step 1)
- Math: Preset and user definable expressions
- Statistics: Histogram View mode: mean, sigma, # of bins and # of samples
- Line frequency detection: Auto detect: 50 Hz or 60 Hz
Front panel operation:
- Front panel interface: 4.3″ TFT color display with keypads and rotary knob
- View mode: Meter view, Graph view, Histogram view and Roll view
- Hard keys: Single Trigger and Run/Stop control, measurement navigation keys, Rotary Knob and Cursors, Ammeter Input and V Source Output control, Cancel/Local
- Soft keys: Function, System and Input Assist keys
- Indicators: Channel (measurement) status, System status
- LEDs:
- Power (color changes when charging)
- Input and Output (color changes when high voltage sourcing)
Input/Output specifications:
- Meter input connector: Three lug triaxial for ammeter on rear panel
- Maximum input:
- Ammeter: 30 mA,
- Voltmeter: 40 V
- Source output connector: Two banana jacks on rear panel
- Maximum common mode voltage: Meter common: 500 V peak: V Source 1,000 V peak
- Isolation (Meter common to chassis: > 10 GΩ, < 500 pF
- Analog output: 2 V for full range input, non-inverting in voltage and cur measurement modes, 1 kΩ output impedance
- Interlock: 2.5 mm pluggable terminal block, 4 pin
- External trigger:
- Trigger input: BNC
- Logic: Programmable edge triggered
- Min pulse width: 10 µs
- Trigger output: BNC
- Logic: Prgrammable edge triggered
- Min pulse width: 10 µs
- Digital I/O:
- Connector type: DSUB female 9 pins
- Input/output pins: DIO 7 pins, +5V, GND
- Absolute max input voltage: 5.25 V
- Absolute min input voltage: -0.25 V
- Max logic L input voltage: -0.8 V, Pull-up to 5 V by 5 kΩ
- Min logic H input voltage: 2.0 V, Pull-up to 5 V by 5 kΩ
- Max source cur: 1 mA @ Vo = 0 V
- Max sink cur: 50 mA @ Vo = 5 V
- 5 V power supply pin: Limited to 500 mA, resettable fuse protected
Voltage range: 2 V
- Display resolution: 1 µV
- Accuracy ±(% + offset): 0.025 + 40 µV
Voltage range: 20 V
- Display resolution: 10 µV
- Accuracy ±(% + offset): 0.025 + 400 µV
Display resolution: 1 µV
Accuracy ±(% + offset): 0.025 + 40 µV
Display resolution: 10 µV
Accuracy ±(% + offset): 0.025 + 400 µV
Ranging: Automatic or manual
Aperture time (integration time): 10 µs to 100 PLC
Offset cancelling: Null, Zero correct
Digital filter: Medium filter (2R + 1, R = 1 to 15), moving average (1 to 100, step 1)
Math: Preset and user definable expressions
Statistics: Histogram View mode: mean, sigma, # of bins and # of samples
Line frequency detection: Auto detect: 50 Hz or 60 Hz
Front panel interface: 4.3″ TFT color display with keypads and rotary knob
View mode: Meter view, Graph view, Histogram view and Roll view
Hard keys: Single Trigger and Run/Stop control, measurement navigation keys, Rotary Knob and Cursors, Ammeter Input and V Source Output control, Cancel/Local
Soft keys: Function, System and Input Assist keys
Indicators: Channel (measurement) status, System status
LEDs:
- Power (color changes when charging)
- Input and Output (color changes when high voltage sourcing)
Power (color changes when charging)
Input and Output (color changes when high voltage sourcing)
Meter input connector: Three lug triaxial for ammeter on rear panel
Maximum input:
- Ammeter: 30 mA,
- Voltmeter: 40 V
Source output connector: Two banana jacks on rear panel
Maximum common mode voltage: Meter common: 500 V peak: V Source 1,000 V peak
Isolation (Meter common to chassis: > 10 GΩ, < 500 pF
Analog output: 2 V for full range input, non-inverting in voltage and cur measurement modes, 1 kΩ output impedance
Interlock: 2.5 mm pluggable terminal block, 4 pin
External trigger:
- Trigger input: BNC
- Logic: Programmable edge triggered
- Min pulse width: 10 µs
- Trigger output: BNC
- Logic: Prgrammable edge triggered
- Min pulse width: 10 µs
Digital I/O:
- Connector type: DSUB female 9 pins
- Input/output pins: DIO 7 pins, +5V, GND
- Absolute max input voltage: 5.25 V
- Absolute min input voltage: -0.25 V
- Max logic L input voltage: -0.8 V, Pull-up to 5 V by 5 kΩ
- Min logic H input voltage: 2.0 V, Pull-up to 5 V by 5 kΩ
- Max source cur: 1 mA @ Vo = 0 V
- Max sink cur: 50 mA @ Vo = 5 V
- 5 V power supply pin: Limited to 500 mA, resettable fuse protected
Ammeter: 30 mA,
Voltmeter: 40 V
Trigger input: BNC
- Logic: Programmable edge triggered
- Min pulse width: 10 µs
Trigger output: BNC
- Logic: Prgrammable edge triggered
- Min pulse width: 10 µs
Logic: Programmable edge triggered
Min pulse width: 10 µs
Logic: Prgrammable edge triggered
Min pulse width: 10 µs
Connector type: DSUB female 9 pins
Input/output pins: DIO 7 pins, +5V, GND
Absolute max input voltage: 5.25 V
Absolute min input voltage: -0.25 V
Max logic L input voltage: -0.8 V, Pull-up to 5 V by 5 kΩ
Min logic H input voltage: 2.0 V, Pull-up to 5 V by 5 kΩ
Max source cur: 1 mA @ Vo = 0 V
Max sink cur: 50 mA @ Vo = 5 V
5 V power supply pin: Limited to 500 mA, resettable fuse protected