The Sorensen SGI600-42 is a Programmable Precision High Power DC Power Supply that is designed for exceptional load transient response, low noise and high power density which is enhanced by a stylish front air intake allowing supplies to be stacked without any required clearance between units. The SGI600-42 combines onboard intelligent controls with outstanding power electronics. These controls enable sophisticated sequencing, constant power mode and save/recall of instrument settings. Looping of sequences makes the SGI600-42 – ideal for repetitive testing. An impressive vacuum fluorescent graphical display in eight languages, context sensitive “soft” keys and front panel keyboard simplify programming of the SGI600-42.Features of the Sorensen SGI600-42 include:
High Power Density: 25kW in a 6U chassis
Wide Voltage Range: 0-600V and 0-42 Amps
Fast Load Transient Response: Protection from undesired voltage excursions
Low RMS and P-P Noise: Suitable for the most sensitive applications
Parallelable up to 150 kW: Expandable as your requirement grows
Modular Design: Upgradeable for the ultimate in investment protection
Sequencing: Program custom waveforms
Easy-to-read Fluorescent Display: Supports English, French, German, Italian, Spanish, Chinese, Japanese, and Korean languages
16-bit Resolution: Optional IEEE-488.2 + RS-232C + Ethernet provides precise control
Ethernet Control: LXI Class C compliant communication through integrated web server
Direct Relay Control: Control output and sense isolation relays, along with polarity relays. (Ethernet Option Only)
Hardware Trigger: Ethernet Option Only
Specifications of the Sorensen SGI600-42 include:Common:
Remote Sense: Load-line loss compensation for models <= 100 V is 10% above full scale voltage total (5% per load-line), and models > 100 V is 4% above full scale voltage total (2% per load-line).
Parallel Operation: Up to 5 units may be paralleled for additional cur within the power supply single-unit specifications, with exception of the DC output cur set accuracy. Additional paralleled SG units will add 0.3% inaccuracy per unit. To parallel more than 5 units,contact factory.
Series Operation: Up to 2 units (see Output Float Voltage)
Input:
Nominal Voltage 3 phase, 3 wire + ground: 208/220 VAC (operating range 187 – 242 VAC); 380/400 VAC (operating range 342 – 440 VAC); and 440/480 VAC (operating range 396 – 528 VAC)
Frequency: 47 – 63Hz (optional 400Hz @ 208VAC, does not carry CE, UL or CSA markings)
Power Factor: >0.9 typical at 208/220 VAC input; >0.78 typical at 380/400 VAC input; and >0.69 typical at 440/480 VAC input
Protection (typical): ½ cycle ride-though on all three phases, 3 cycle ride through on single phase; missing phase shutdown (800V model 6.4 msec on all 3 phases)
Environmental:
Operating Temperature: 0 to 50º C
Storage Temperature: -25º C to 65º C
Humidity Range: Relative humidity up to 95% non-condensing, 0º C – 50º C
Altitude: Operating full power available up to 5,000 ft. (~1,500 m), derate 10% of full power for every 1,000 feet higher; non-operating to 40,000 ft. (~12,000 m)
Cooling: Front and side air inlet, rear exhaust. Temperature controlled, variable speed fans. Units may be stacked without spacing.
Regulatory: Certified to UL/CSA 61010 and IEC/EN 61010-1, CE Compliant, Semi-F47 Compliant
Front Panel Dust Filter: 30 PPI (Pores Per Inch) – must ensure adequate airflow and / or derate max. temperature. 3U unit only.
Physical:
Dimensions: Width: 19.00” (48.3 cm), Depth 25.0” (63.5 cm) Height: 5-15 kW units: 3U – 5.25” rack mount (13.34 cm) 20-30 kW units: 6U – 10.5” rack mount (26.67 cm)
Weight: 3U < 80 lbs. (36 kg) 6U
Output:
Ripple & Noise (Voltage Mode, Typical): Ripple and noise specified at full load, nominal AC input. Noise measured with 6ft. cable, 1µf at load
Ripple (Cur Mode): <+/- 0.04% of full scale rms cur
DC Voltage Slew Rate: 45A / ms typical – resistive load
Line Regulation (with sense wires used): (±10% of nominal AC input, constant load); Voltage Mode: +/- 0.01% of full scale; and Cur Mode: +/- 0.05% of full scale
Load Regulation (with sense wires used): no load to full load, nominal AC input); Voltage Mode: +/- 0.02% of full scale; and Cur Mode: +/- 0.1% of full scale
Load Transient Response: Recovers within 1ms to +/-0.75% of full-scale of steadystate output for a 50% to 100% or 100% to 50% load change
Efficiency: 87% typical at nominal line and max load
Stability: ±0.05% of set point after 30 minute warm-up and over 8 hours at fixed line, load and temperature
Temperature Coefficient: 0.02%/ C of maximum output voltage rating for voltage set point 0.03%/ C of maximum output cur rating for cur set point
Output Float Voltage: Negative terminal within +/- 300 V of chassis potential. (We recommend the use of optional isolated analog Interface (IAI).) Supplies in “series” should be the same output voltage/cur.
Programming Specifications of the Sorensen SGI600-42 include:Front Panel Display:
Voltage Accuracy: +/- 0.1% of full scale
Cur Accuracy: +/- 0.4% of full scale
Resolution: 4.0 digits
Remote Analog Interface:
Voltage Accuracy: +/-0.25% of full scale for 0-5 V range, +/-0.5% of fullscale for 0-10 V range
Cur Accuracy: 0.8% of full scale
Remote Digital Interface:
Voltage Accuracy: +/- 0.1% of full scale
Cur Accuracy: +/- 0.4% of full scale
Resolution: +/-0.002% of full scale
OVP:
Accuracy: +/- 1% of full scale
Resolution: +/-0.002% of full scale
User I/O:
Disconnect & Polarity-reversal relay control (Only available with Ethernet Option)
Read-Back / Monitoring Specifications of the Sorensen SGI600-42 include:Front Panel Display:
Voltage Accuracy: +/- 0.1% of full scale
Cur Accuracy: +/- 0.4% of full scale
Resolution: 4.0 digits
Remote Analog Interface:
Accuracy: +/-1.0% of full scale (0 – 10V)
Remote Digital Interface:
Voltage Accuracy: +/- 0.1% of full scale
Cur Accuracy: +/- 0.4% of full scale
Resolution: +/-0.002% of full scale
User I/O:
Disconnect & Polarity-reversal relay control (Only available with Ethernet Option)